Design considerations for particulate instrumentation employing laser light scattering
This research is concerned with the design and development of a light scattering instrument that performs in situ and on-line particle density distribution measurements of polydisperse aerosols. The particles of principal interest are water droplets having diameter 25 < d < l000 μm. A pulsed GaAs laser is used as the source and the light detection system is built around solid state detectors and electronics. The scattering theory of Mie is applied to the design of the light scattering instrument.
The principal objectives of this system are to present an optically defined, external sampling volume of the order of one cubic centimeter to a particulate-laden air flow and to realize an instrument response function such that the density distribution of an ensemble of particles can be accurately determined. The approach to these objectives was to investigate off-axis and in-line optical configurations, each having distinct merits depending upon the application. It was found from interpretation of the Mie theory that off-axis systems can be optimally designed for detection of drift particles in a fluctuating ensemble of background particles and that in-line systems operating near the forward lobe collect more radiation with a correspondingly smaller minimum detectable particle.
Prototype systems were constructed and measured response characteristics showed good agreement with the theory. Problems encountered with the large, external scattering volume were evident in the form of non-ideal response functions because of inhomogeneities in the spatial intensity distribution of the laser source and variable collection of scattered light, also dependent upon position in the sampling volume. These difficulties were solved by introduction of a light integrating optical system and by proper design of receiver optics to minimize collection or solid angle effects.
Considerable experimentation with the prototype instrumentation has shown that the design parameters are highly interrelated. The salient characteristics of the final system are:
1. A peaking response function for a given size particle resulting in particle size determination of +l0% with 68% certainty.
2. An external sampling volume of several cubic centimeters while maintaining peaking response.
3. Relationship between average instrument voltage response v and particle diameter d of v = Kd2 in accordance with theory.
Finally, it was shown that typical particle size distributions can be recovered by appropriate mathematical techniques, even for instruments exhibiting non-ideal response characteristics.