Use of a tunable laser to increase the dynamic measurement range of copper concentration by absorption spectroscopy
Atomic and molecular absorption spectroscopy is presently limited in measurable concentration range due to, among other things, the inability of instrumentation to provide high intensity, highly monochromatic incident light for the measurement of high absorbances. A tunable laser can provide such light at the wavelength of maximum sample absorbance and, thus, an experiment was performed to determine the advantages of using a tunable laser as a light source in the absorption colorimetric analysis of copper concentration.
The best available tunable laser to be chosen was a pulsed laser, and appropriate detection electronics were designed to quantitatively measure the light intensities. A double beam system was used to correct for shot-to-shot laser intensity variations. Standard samples of known concentration were prepared and their absorbances measured using the developed system. A curve which was least-squares fitted to the concentration-versus-optical density data proved to be linear to a concentration of 15.3 μg/ml, at which point it became nonlinear due, apparently, to chemical effects in the solution. This value represents an apparent increase in the linearity range of a factor of four over that previously obtained using conventional equipment for this complex.