Characterization of Advanced Detector Technologies for Neutron Science Applications
Date Issued
August 2026
Author(s)
Hoegberg, Erick K
The University of Tennessee
Advisor(s)
Eric Lukosi
Additional Advisor(s)
Jason Hayward
Mathieu Benoit
Abstract
Neutron science is a set of experimental methodologies that rely on the unique properties of neutrons to obtain new information about an object, process, material or fundamental physics. Many scientific facilities exist around the world to provide neutrons for these experiments which have wide application in many fields. The cutting edge of neutron science is limited by the performance of the neutron detectors themselves, improvement of which allows experiments to be performed with more accuracy, revealing more information. This work contains an analysis of several detector technologies, enabled by the timepix3 ASIC, designed to enhance the scientific capabilities of neutron scattering facilities, particularly those at HFIR and SNS at ORNL.
In part one, suitability of the novel Lithium-containing semiconductor Lithium Indium Phosphor Selenide (6LiInP2Se6, LIPSe) is evaluated. This material offers good charge-carrier properties and could be mated with a timepix or other detector framework to provide a neutron detector with excellent timing and spatial resolution with several advantages over existing technologies in the same niche. Firstly, an analysis is presented on the change in charge carrier properties in variants doped with Cadmium making them easier to grow at scale, concluding that charge carrier properties are retained in this variant. Secondly, the radiation tolerance of this material was evaluated and found suitable to survive the high-intensity radiation environment present at neutron beam lines up to a fluence representing approximately one year of operational time, indicating its potential utility in the field and ability to be economically integrated into regular maintenance cycles.
In part two, detection schemes based on the TPX3Cam system, a light-sensitive Timepix3 detector are investigated. These rely on a neutron-sensitive light source to provide input to an image intensifier which is read out by the timepix, providing images over a large area which are well-resolved in space and time. Topics investigated include the gamma discrimination capabilities when paired with several types of neutron sensitive scintillators and a comparison between these and large-area MCPs, and investigations into some artifacts encountered. The advantages and limitations of each configuration are discussed alongside conclusions about the best use of each one.
Degree
Doctor of Philosophy
Major
Nuclear Engineering
File(s)![Thumbnail Image]()
Name
HOEGBERG-PRIMARY-2026.pdf
Size
4.42 MB
Format
Adobe PDF
Checksum (MD5)
009af8a300feb6a29ba51cf61166aa47