Determination of the Muon Lifetime With a Water Cherenkov Detector
Date Issued
August 2026
Author(s)
The University of Tennessee
Advisor(s)
Stefan Spanier
Additional Advisor(s)
Miguel Madurga Flores
Marianne Breinig
Christine Nattrass
Abstract
An experimental setup for measuring the charge-averaged muon lifetime with a water Cherenkov detector, using a high-speed digitizer to read out a photomultiplier tube, has been developed for an advanced student laboratory. An approximately 0.40 m3 cylindrical stainless-steel container has been constructed at the University of Tennessee Department of Physics and Astronomy and equipped with a KamLAND type photomultiplier tube to detect Cherenkov light produced by cosmic muons in water. A digitizer-based readout continuously captures waveform data. An algorithm identifies two pulses separated by more than 100 ns, associated with a primary negative or positive muon that stops in the water and subsequently decays into an electron or positron, respectively. A total of about 730k decay candidate events were collected continuously over 22 days. The mean muon lifetime was determined from a subset to be τ [tau] = 2.12 ± 0.12 µs [mu], with a systematic uncertainty of 0.03 µs, in comparison to an expected value of τ = 2.04±0.01 µs based on the world average muon lifetime and an assumed charge ratio.