Analysis of cigarette smoke by fourier transform infrared spectroscopy
The application of Fourier transform infrared spectroscopy to the quantitative analysis of several components in whole dilute tobacco smoke was demonstrated. The absorption cell which had a path length of 18 cm was part of a cigarette smoking system similar to the intermittent inhalation exposure devices used in smoking and health research with rodents.
Concentrations were measured for carbon monoxide, carbon dioxide, methane, ethylene, and methanol in 7 to 22% smoke. The precision (relative standard deviation) of a single observation in 22% smoke ranged from 3-4% for carbon monoxide, carbon dioxide, and methane (best cases) to 34% (worst case) for ethylene.
Absorbances measured for isoprene and hydrogen cyanide followed known concentrations in different cigarette smokes.
It was shown that the concentrations of each of these constituents are approximately constant during the 30-sec puff stand time. The results obtained show clearly the possibility of determining several smoke components in a single puff by means of FT-IR spectroscopy.