A continuous flow centrifugal separator for use in immunodepletion : theory, design, and experiment observations
A continuous flow centrifugal separator was developed that is capable of separating and collecting fluid plasma and packed cellular components of whole blood on a continuous ex vivo basis. Whole blood is continuously fed into the separator, separated into plasma and cells, and removed from the separator. This process is one step in a rabbit immunodepletion system in which the procurement of cell free plasma is of primary importance. After removal from the separator, plasma can be routed to the next step of the system, depleted of antibodies or antigens, and then remixed with the blood cells for reinjection into the live rabbit. Because it is not necessary to alter either the flow into or out of the centrifuge or to alter the centrifugal force field while filling or collecting, large volumes of whole blood can be processed rapidly in a small volume centrifuge. The continuous flow method represents a six-fold increase in antibody or antigen production over conventional methods.
Sedimentation was chosen as the most expedient method of separating cells from plasma. Centrifugal force accelerates this naturally occurring process and therefore provides an efficient method of blood separation. Even utilizing this method, cell free plasma is difficult to obtain because of the relatively slow sedimentation rate of platelets. Theoretical sedimentation laws were used to identify the critical parameters of plasma isolation. Stokes' law, with a correction for shape and concentration, provided the method of calculating the platelet sedimentation velocity in plasma.
The actual configuration of the centrifuge bowl was based on the axial flow class of centrifugal blood separators. The design allows incorporation of the bowl into a drive unit based upon Adams' anti-twister mechanism as developed by the American Instrument Company.
Because of the difficulty of testing with a live rabbit, evaluation of the bowl's ability to separate cells efficiently from plasma was based on data obtained using freshly drawn human blood. Experimental runs were made in Silver Spring, Maryland at the American Instrument Company. All the erythrocytes and leucocytes were removed, but a small percentage of platelets (2% - 3%) remained in the plasma line. However, a 97% - 98% removal efficiency is considered sufficient for the immunodepletion procedure. The small number of platelets left in the plasma as it flows through the cartridge are not expected to interfere with the depletion of antibodies or antigens. When the rotor was operating efficiently and separating blood components effectively, decreases in rotational speeds or increases in plasma pumping rate for given whole blood feed rates allowed platelets as well as erythrocytes and leucocytes to exit from the plasma port. Even slight changes in these parameters caused significant changes in cell removal efficiencies. Exceedingly careful and accurate design calculations are an absolute necessity to design, machine, and obtain a rotor which will separate blood properly into desired fractions.
These experiments proved the Davidson-designed centrifugal rotor to be an effective and practical tool for use in rabbit immunodepletion procedures. The rotor bowl is completely closed, allows no dilution of blood, creates no biological problems such as clotting or excessive cell damage, has an acceptable internal blood volume, operates at acceptable flow rates, and efficiently separates cellular components from plasma. This precision device could be enormously useful for certain biological investigations exploring the immune response of the mammalian system. In addition, the rotor could be adapted for use in platelet concentrated plasma procurement or any other two-phase separation of whole blood. Using the design criteria herein established this rotor/bowl could be proportionately scaled up for use with larger mammals.