Detailed analysis of the calibration of spectra ; an analysis of nu
The first part of this work is a detailed analysis of the calibration of spectra by inserting standard calibration lines in the spectra at regular intervals. The two major types of error, the measurement error and the interpolation error, are examined. The concept of the proper density of the calibration lines is introduced. Seven scans of CO spectra are used to examine the reproducibility and the proper density of the calibration lines of the 5-meter spectrometer. In addition, a comparison is made of two different methods to handle data of several repeated scans.
In the second part of this work the ν9 + ν12 near 2024 cm-1 is recorded using the 5-meter Littrow instrument. The resolution of the spectra is about 0.020 cm-1 Both the Eu and A2u components of the Eu x Eg type vibration are observed. Ground state combination differences have been calculated and the ground state molecular parameters B0, DJ0 , DJK0 are found to be 0.66308141 (± 0.190 x 10-4) cm-1, 1.134 x 10-6 (±0.272 x 10-7) cm-1, and 2.396 x 10-6 (±0.427 x 10-6) cm-1 , respectively. Various perturbations are observed but not analyzed.