Repository logo
Log In(current)
  1. Home
  2. Colleges & Schools
  3. Graduate School
  4. Doctoral Dissertations
  5. Precision rectifier detectors for AC resistance bridge measurements with application to temperature control systems for irradiation creep experiments
Details

Precision rectifier detectors for AC resistance bridge measurements with application to temperature control systems for irradiation creep experiments

Date Issued
March 1997
Author(s)
Duncan, Michael Gary
Advisor(s)
Graham W. Hoffman
Additional Advisor(s)
Marshall Pace
Robert E. Bodenheimer
J. P. Tillman
T. Vaughn Blalock
Permanent URI
https://trace.tennessee.edu/handle/20.500.14382/60030
Abstract

The suitability of several temperature measurement schemes for an irradiation creep experiment is examined. It is found that the specimen resistance can be used to measure and control the sample temperature if compensated for resistance drift due to radiation and annealing effects. A modified Kelvin bridge is presented that allows compensation for resistance drift by periodically checking the sample resistance at a controlled ambient temperature.

 A new phase-insensitive method for detecting the bridge error signals is presented. The phase-insensitive detector is formed by averaging the magnitude of two bridge voltages. Although this method is substantially less sensitive to stray reactances in the bridge than conventional phase-sensitive detectors, it is sensitive to gain stability and linearity of the rectifier circuits. Accuracy limitations of rectifier circuits are examined both theoretically and experimentally in great detail. Both hand analyses and computer simulations of rectifier errors are presented. 

 Finally, the design of a temperature control system based on sample resistance measurement is presented. The prototype system is shown to control a 316 stainless steel sample to within a 0.15°C short term (10 sec) and a 0.03°C long term (10 min) standard deviation at temperatures between 150 and 700°C. The phase-insensitive detector typically contributes less than 10 ppm peak resistance measurement error (0.04°C at 700 C for 316 stainless steel or 0.005°C at 150°C for zirconium).

 
Degree
Doctor of Philosophy
Major
Electrical Engineering
File(s)
Thumbnail Image
Name

Thesis77b.D852.pdf

Size

27.46 MB

Format

Adobe PDF

Checksum (MD5)

283c8d402ca21d6dfe1b07cc9f0ae128


University Libraries

1015 Volunteer Boulevard
Knoxville, TN 37996
865-974-4351

Map & Directions
Donate to the Libraries
  • About
  • John C. Hodges Society
  • Speaking Volumes magazine
  • Outreach
  • Directory
  • Employment
  • Policies
  • Library Intranet
University of Tennessee power T logo

The University of Tennessee, Knoxville
Knoxville, Tennessee 37996
865-974-1000

Events
A-Z
Apply
Privacy
Map
Directory
Give to UT
Accessibility

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science