The effect of flow parameters on dilatations in stenotic arterial models
A sudden constriction or narrowing of the cross-sectional area of a blood vessel is referred to as a stenosis. These constrictions are usually caused by intravascular lesions. One common adverse effect of a severe stenosis is a weakening and bulging of the vessel wall distal to the stenosis, commonly referred to as poststenotic dilatation. Such a condition may lead to complete structural fatigue of the vessel wall, permitting the vessel to rupture, and, in some cases, causing death.
The purpose of this research project was to conduct a model study of stenotic vessels under various flow conditions, and to determine, in vitro, the conditions to produce poststenotic dilatation. The long-range goal is to compare such studies with rigid model studies and in vivo dilatation, in hopes of determining a prognostic tool to determine the likelihood of poststenotic dilatation occurrence in patients' stenotic vessels.
Latex models of a stenosed canine femoral artery were fabricated, and their moduli of elasticity evaluated. Utilizing dimensional analysis and similitude, pertinent dimensionless terms were developed which could be reproduced in vitro. A series of tests were run using small models (0.6 cm diameter) in a mock circulatory system. Other tests were conducted with larger models (2 cm diameter) in a different system. Most of the models had an area reduction of 75% at the throat of the stenosis.
Under modeled normal physiological conditions, poststenotic dilatation did not occur in any of the models. Rather, the models developed prestenotic dilatations. Increasing the mean velocity to abnormally high values only accelerated prestenotic dilatation. However, when the modeled pulse rate was increased above its normal value and all other variables were held at normal values, poststenotic dilatation did develop in the larger models. Unlike the larger models which were made entirely of latex rubber, the stenoses in the smaller models were machined from Plexiglas, which may have contributed to the prestenotic dilatations in these models. The only time poststenotic dilatation developed in a small model was when the support for the model distal to the stenosis was partially removed. Pressure variations along the inside wall of the models were obtained by a specially designed pressure catheter. The results indicated no localized high pressure region within the tube.
From the results of these experiments, it is believed that mechanical effects are complemented by biological effects to produce poststenotic dilatation. Furthermore, the significance of pulse rate in their production has been indicated. Other important parameters appear to be the method of support, the material properties of the model, and the degree of constriction. Further studies are needed to determine more information on this complicated but important problem.