The reproducibility of oxygen uptake and heart rate responses at the same work rate on different bicycle ergometers
The purpose of this study was to determine the reproducibility of steady state oxygen uptake (VO2) and heart rate responses for the same subject on different bicycle ergometers which were calibrated in an identical manner. Four subjects exercised on each of three bicycle ergometers during which VO2 and heart rate responses were measured at two different workloads (88 watts and 147 watts). Each subject was tested twice at each workload. Test and retest VO2 and mean heart rate values obtained at specific workloads were subjected to a two-way analysis of variance. There were no differences (P > 0.05) among bicycle ergometers at the low or high workloads for VO2. Mean heart rate values for the low workload were not significantly different (P > 0.05) from each other. However, heart rate values were significantly different (P < 0.05) among bicycle ergometers at the high workload.