Solution microcalorimeter for measuring heats of solution of radioactive elements and compounds
An isoperibol, solution microcalorimeter capable of measuring the heats of solution of submilligram quantities of radioactive elements and compounds was constructed. In particular the calorimeter would be used to measure the heats of solution of some transplutonium elements (and compounds) in order to determine the heat of formation of the element's aquo-ion.
The microcalorimeter vessel is constructed of tantalum metal, with a nominal volume of 5 cm3. Its energy equivalent is 24 J K-1 when containing 5 cm3 H2O. The thermal leakage modulus is 0.010 min-1 A thermistor is employed as the temperature sensor. The operating sensitivity is about 1 X 10-5 K (300 μJ).
A novel sample container was developed. It is fabricated from Pyrex, and its basic design is a barrel shape with thin glass membranes sealed across the two open ends of the container. Sample containers were subjected to extensive leak tests with no leaks being found in the seals or membranes. The chemical reaction is initiated by puncturing the membranes with a sharp wire.
The facility has the capability of a data acquisition/analysis mode of operation. The output of the system is digitized with a digital voltmeter, and these data are stored on magnetic tape. The voltages necessary to determine the power delivered to the calorimeter during calibration heatings are measured automatically. They along with the bridge resistance settings are also stored on magnetic tape. The operation is controlled by a small, in-house computer. The stored data are then analyzed by the computer.
The performance of the calorimetry system was tested by measuring the heat of solution of NBS Standard Reference Material 724a, which is tris(hydroxymethyl)aminomethane (TRIS), and magnesium metal. The results of the TRIS experiments, at a concentration of 1 g dm-3 in 0.1 N HCl at 298 K, yielded a heat of solution of -29.606 ± 0.063 kJ mol-1. The magnesium experiments, in 1 N HCl at 298 K, gave a heat of solution of -465.965 ± 1.136 kJ mol-1. The results of these experiments demonstrate that the (present) sample containment and stirring configurations are especially appropriate for experiments with metals and somewhat less suitable for powdered samples.
The heat of solution of curium-248 metal in 1 N HCl at 298 K was measured. The experiments, which should not be considered definitive, yielded a heat of solution of -606.4 ± 1.8 kJ mol-1. A single measurement in 6 N HCl gave a heat of solution of -602.3 kJ mol-1. From these results the heat of formation of Cm3+ (aq) is calculated to be -607.2 ± 2.5 kJ mol-1.
Thesis78b.R382.pdf
4.94 MB
Adobe PDF
4f2562b6e148a8ff86495ea1ed9d3846