Evaluation of Stent Retriever Performance and Feasibility of an Auxetic Stent Retriever Design
Date Issued
May 2026
Author(s)
Poulos, Demitria Alexa
The University of Tennessee
Advisor(s)
Bryan Good
Additional Advisor(s)
Colleen Crouch
Caleb Rucker
Michael Froehler
Abstract
An ischemic stroke, the result of a blood clot occluding a cerebral vessel, accounts for 87% of all strokes. More specifically, acute ischemic stroke due to large vessel occlusion accounts for 90% of 6-month post stroke mortality. The key to achieving good neurologic outcome in stroke patients is to rapidly restore blood flow, best achieved through a mechanical thrombectomy (MT). During stent retriever (SR) MT, a SR is deployed across the occluding blood clot and the entire system is retracted to restore flow. Studies have shown that MT is more likely to fail in complex occlusion locations; however, data relating to SR performance in these locations is lacking.
This project focuses on investigating SR behaviors and forces during SR MT that directly influence clot dislodgement in tortuous segments. In vitro thrombectomies were performed in a flow loop with models of increasing complexity to measure the SR removal force (i.e. force to dislodge the device from the occlusion site) and understand the influence of occlusion location on retrieval force. A 2D arterial model mimicking the tortuosity of an internal carotid/middle cerebral artery was developed and the deformations of SRs at deployment and retrieval in the model were quantified. To further understand SR behavior at deployment, the outward radial forces exerted by SRs in tortuous geometries were measured. Once an improved understanding of SR behavior was achieved, embolus analogs (EAs) were injected into the tortuous model to mimic stroke and MTs were performed with SRs of varying geometry and type.
This study determined that SRs experience collapse, limiting contact with the model and EA, at both deployment and retrieval and that tortuosity leads to significant increases in collapse and decreases in SR outward forces. To minimize collapse in these complex geometries, the feasibility of applying auxetic concepts to SRs was evaluated. Through computational and preliminary studies, a non-auxetic SR was shown to decrease in diameter while auxetic SRs expanded under tension and resisted significant collapse under bending. These results indicate the potential for applying auxetics to SR MT and their potential to increase first pass effect rates in stroke patients.
Degree
Doctor of Philosophy
Major
Biomedical Engineering