Read-out electronics for a single photon emission computed tomography (SPECT) small animal imaging system
Read-out electronics have been designed for a new single photon emission computed tomography (SPECT) system. The SPECT system was incorporated into the existing Oak Ridge National Laboratory's MicroCAT X-ray CT imaging system for the purpose of complete dual-modality small animal imaging. Data acquisition software for SPECT has been written using the Labwindows CVI software package by National Instruments. An initial board-level design of the read-out electronics with discrete parts and 4 analog channels has been implemented to characterize the detector and acquire initial images. Then, a prototype application specific integrated circuit (ASIC) with a single analog channel was developed in the AMI 1.6µm process through the MOSIS microelectronics prototyping program. Theoretical, simulated, and measured results of the analog channel are discussed in detail. The analog channel was found to have a DC gain of 138dB, a time-to-peak of 8.8µs, a total output noise voltage of 0.8mVRMS, and 17.5mW of power dissipation.
Thesis2001S95.pdf
21.04 MB
Adobe PDF
73d1f2c88ec13437f47004206b36a9fe