Design of a Robotic Instrument Manipulator for Endoscopic Deployment

Date Issued
August 1, 2017
Author(s)
Ponten, Ryan  
Advisor(s)
Daniel C. Rucker
Additional Advisor(s)
William Hamel
Jindong Tan
Permanent URI
https://trace.tennessee.edu/handle/20.500.14382/41106
Abstract

This thesis describes the initial design process for an application of continuum robotics to endoscopic surgical procedures, specifically dissection of the colon. We first introduce the long-term vision for a benchtop dual-instrument endoscopic system with intuitive haptic controllers and then narrow our focus to the design and testing of the instrument manipulator itself, which must be actuated through the long, winding channel of a standard colonoscope.


Based on design requirements for a target procedure, we analyze simulations of two types of continuum robots using recently established kinematic and mechanic modeling approaches: the concentric-tube robot (CTR) and the concentric agonist-antagonist robot (CAAR). In addition, we investigate solutions to the primary engineering challenge to this system, which is accurately transmitting joint motion through exible, hollow shafts. Based on our study of the manipulator simulations and transmission shafts, we select instrument designs for prototyping and testing. We present approaches for controlling the position of the robotic instrument in real-time using an input device, and demonstrate the degree of control we can achieve in various configurations by performing time trial experiments with our prototype robotic instruments. Our observations of the manipulator during testing inform us of sources of error, and we conclude this report with suggestions for future work, including shaft design and alternative continuum manipulator approaches.

Subjects

manipulator

workspace

endoscopic instrument...

degrees of freedom

dexterity

Disciplines
Applied Mechanics
Biomedical Devices and Instrumentation
Computer-Aided Engineering and Design
Electro-Mechanical Systems
Degree
Master of Science
Major
Mechanical Engineering
Embargo Date
August 15, 2018
File(s)
Thumbnail Image
Name

ThesisRyanPonten.pdf

Size

22.29 MB

Format

Adobe PDF

Checksum (MD5)

82581df693fdd3930da5d7fa73e8203d


University Libraries

1015 Volunteer Boulevard
Knoxville, TN 37996
865-974-4351

Map & Directions
Donate to the Libraries
  • About
  • John C. Hodges Society
  • Speaking Volumes magazine
  • Outreach
  • Directory
  • Employment
  • Policies
  • Library Intranet
University of Tennessee power T logo

The University of Tennessee, Knoxville
Knoxville, Tennessee 37996
865-974-1000

Events
A-Z
Apply
Privacy
Map
Directory
Give to UT
Accessibility

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science