Increasing Hydrophilicity and Transparency of Diamond-Like Carbon Thin Films with Dopants for an Anti-Fogging Laparoscope Coating

Date Issued
December 1, 2021
Author(s)
Bull, Anna  
Advisor(s)
Jacqueline A. Johnson
Additional Advisor(s)
Jacqueline A. Johnson, Russell L. Leonard, Trevor M. Moeller
Abstract

Laparoscopes are prone to fogging which can lead to a limited field of view during surgical procedures. Current methods of mitigating fogging issues are not efficient or can require costly modification to the laparoscope. Previous studies of diamond-like carbon (DLC) coatings found doping the films improved hydrophilic qualities, suggesting their possible use as antifogging coatings for laparoscopes. For this work, two series of DLC films, doped with either SiO and Al2O3 were investigated.


The biocompatibility, transparency, and stability of these films were assessed through cellular assays, spectrophotometry, and simulated body fluid soaking experiments. Contact angle and surface energy measurements were performed to assess the hydrophilic qualities of the films. Plasma cleaning was utilized as a surface treatment to improve hydrophilicity; time-studies were performed to assess the stability of this treatment. Raman spectroscopy, atomic force microscopy, and adhesion tests were performed to assess the physical characteristics of the films.

The silicon monoxide doped films demonstrated improved transparency and hydrophilic qualities. Samples subjected to plasma cleaning had contact angles under 5° when measured within 60 minutes of treatment. From the time-studies performed, the hydrophilicity of the SiO doped films was improved for over 24 hours after treatment. The aluminum oxide doped films demonstrated transparency results of up to 97% over the visible spectrum indicating this dopant helped improve transparency. After plasma cleaning samples demonstrated a contact angle of 8° within 60 minutes of treatment. Both series demonstrated similar biocompatibility results with cellular assay viability values being statistically similar to the control media. No delamination of the films from either series was observed when soaked in simulated body fluid over the course of 40 weeks. The results of this research show promise for DLC as an antifogging coating for laparoscopes.

Subjects

Diamond-like carbon

Pulsed Laser Depositi...

Antifogging

Disciplines
Biomedical Engineering and Bioengineering
Degree
Master of Science
Major
Biomedical Engineering
File(s)
Thumbnail Image
Name

ThesisFinal_Bull.pdf

Size

1.88 MB

Format

Adobe PDF

Checksum (MD5)

10128e196d09190a83efda58737bd849

Thumbnail Image
Name

Thesis_Trace_09212021.docx

Size

10.76 MB

Format

Microsoft Word XML

Checksum (MD5)

d5013d9464399008d45f8613de8b8a17

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Privacy policy
  • End User Agreement
  • Send Feedback
  • Contact
  • Libraries at University of Tennessee, Knoxville
Repository logo COAR Notify