Repository logo
Log In(current)
  1. Home
  2. Colleges & Schools
  3. Graduate School
  4. Masters Theses
  5. The effect of acrylic and PVC earmold material on REOR and REAR
Details

The effect of acrylic and PVC earmold material on REOR and REAR

Date Issued
May 1, 1999
Author(s)
McIntosh, Kimberly Jon
Advisor(s)
Samuel B. Burchfield
Additional Advisor(s)
James Thelin, Mark Hedrick
Permanent URI
https://trace.tennessee.edu/handle/20.500.14382/46583
Abstract

Experimental earmolds which were otherwise identical but constructed from acrylic and polyvinylchloride material were compared by REAR and found to produce difference amplified responses in the ear canals of the children that participated in the study. The acrylic material produced uniformly and significantly greater amplitude. Pre- and post-test hearing aid and real-ear reliability results suggest that the observed REAR differences were not produced by measurement error and that they were not produced by insertion loss differences. This means that the observed response differences was produced by the transmission properties of the earmold material. The implication of this finding is that soft earmolds, which are often specified to prevent external feedback, may be compromising the acoustic response of the children's hearing aids.


The concept of fitting children with soft attenuating earmolds despite decreased amplitude in the hearing aid response, as well as the employment of acrylic earmolds in pediatric populations with mild-to-moderate hearing losses is advanced.

Degree
Master of Arts
Major
Audiology
File(s)
Thumbnail Image
Name

Thesis99.M254.pdf

Size

590.16 KB

Format

Unknown

Checksum (MD5)

411bfbfd920e6e2deed4363e1ae6b19f

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science

  • Privacy policy
  • End User Agreement
  • Send Feedback
  • Contact
  • Libraries at University of Tennessee, Knoxville
Repository logo COAR Notify