  6. Accuracy of the Cosmed K5 portable calorimeter
Details

Accuracy of the Cosmed K5 portable calorimeter

Date Issued
January 1, 2019
Author(s)
Crouter, Scott E.  
LaMunion, Samuel R.  
Hibbing, Paul R.
Kaplan, Andrew S.
Bassett Jr., David R.
Permanent URI
https://trace.tennessee.edu/handle/20.500.14382/17381
Abstract

Purpose

The purpose of this study was to assess the accuracy of the Cosmed K5 portable metabolic system dynamic mixing chamber (MC) and breath-by-breath (BxB) modes against the criterion Douglas bag (DB) method.

Methods

Fifteen participants (mean age±SD, 30.6±7.4 yrs) had their metabolic variables measured at rest and during cycling at 50, 100, 150, 200, and 250W. During each stage, participants were connected to the first respiratory gas collection method (randomized) for the first four minutes to reach steady state, followed by 3-min (or 5-min for DB) collection periods for the resting condition, and 2-min collection periods for all cycling intensities. Collection periods for the second and third methods were preceded by a washout of 1–3 min. Repeated measures ANOVAs were used to compare metabolic variables measured by each method, for seated rest and each cycling work rate.

Results

For ventilation (VE) and oxygen uptake (VO2), the K5 MC and BxB modes were within 2.1 l/min (VE) and 0.08 l/min (VO2) of the DB (p≥0.05). Compared to DB values, carbon dioxide production (VCO2) was significantly underestimated by the K5 BxB mode at work rates ≥150W by 0.12–0.31 l/min (p<0.05). K5 MC and BxB respiratory exchange ratio values were significantly lower than DB at cycling work rates ≥100W by 0.03–0.08 (p<0.05).

Conclusion

Compared to the DB method, the K5 MC and BxB modes are acceptable for measuring VE and VO2 across a wide range of cycling intensities. Both K5 modes provided comparable values to each other.

Submission Type
Publisher's Version
File(s)
Thumbnail Image
Name

pone.0226290.pdf

Size

1.14 MB

Format

Adobe PDF

Checksum (MD5)

9e26893589b68e0d02d2bce78e706968


University Libraries

1015 Volunteer Boulevard
Knoxville, TN 37996
865-974-4351

Map & Directions
Donate to the Libraries
  • About
  • John C. Hodges Society
  • Speaking Volumes magazine
  • Outreach
  • Directory
  • Employment
  • Policies
  • Library Intranet
University of Tennessee power T logo

The University of Tennessee, Knoxville
Knoxville, Tennessee 37996
865-974-1000

Events
A-Z
Apply
Privacy
Map
Directory
Give to UT
Accessibility

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science